Questions & Answers

The NextSeq Online Community

Join other NextSeq owners in the NextSeq Online Community. Collaborate with Illumina moderators and NextSeq owners. Discuss best practices, troubleshoot, and learn about how others are using NextSeq preparation kits, push-button sequencing, and automated data analysis to fuel their research.

Join Now

Expand All

General

  • What power connections are required for the NextSeq system?

    The NextSeq instrument comes with a region-specific power cord. For more information, see the NextSeq System Site Prep Guide.

  • What are the dimensions of the NextSeq system?

    For instrument specifications, see the Site Prep/Lab Environment page.

Primers

  • Can custom primers be used on the NextSeq system?

    Yes. There are three reservoirs on the NextSeq reagent cartridge reserved for custom primers: #7 for a custom Read 1 primer, #8 for a custom Read 2 primer, and #9 for a custom Index 1 primer or custom Index 2 primer.

    Using a custom Index 2 primer requires the NextSeq 500/550 Kit v2 and NCS v1.4. Using a custom Index 2 primer is not possible with previous kit or control software versions.

Libraries

  • What is the library loading volume and concentration required for sequencing on the NextSeq system?

    Using NextSeq Control Software v1.3, or later, the loading volume is 1.3 ml and the loading concentration is 1.8 pM.

    For more information, see NextSeq System Denature and Dilute Libraries Guide (15048776).

  • How many different libraries can be sequenced on a NextSeq flow cell?

    Libraries are transferred onto the flow cell from a single reservoir on the NextSeq reagent cartridge, reservoir #10. Therefore, any libraries that can be pooled can be sequenced together on the flow cell.

  • What are the supported insert sizes on the NextSeq system?

    Sequencing an insert size of 550 bp is supported with NextSeq v2 reagents. With NextSeq v1 reagents, an insert size of 350 bp is supported.

NextSeq v2.5 Reagent Kits

How does the NextSeq v2.5 flow cell provide improved performance?

The NextSeq v2.5 flow cells use an updated FC chemistry that improves stability and robustness.  This chemistry allows us to ship the flow cells at ambient temperature; once received, store the flow cells at 2°C to 8°C.

Is a software update required to use the NextSeq v2.5 Reagent Kits?

Yes, NextSeq Control Software - NCS v2.2.0 or later is required to run the NextSeq v2.5 Reagent Kits.

Can the NextSeq v2.5 Kits be run on the NextSeq 550Dx instrument?

No, the NextSeq 550Dx instruments require NCS 3.0. which is not compatible with the NextSeq v2.5 Kits.

Is the software customer installable?

Yes NCS 2.2.0 is customer installable.  NCS 2.2.0 will be pushed to registered BaseSpace users, if configured for automatic updates.  Otherwise,  download the software from the Downloads page.

Is there any visible change to the NextSeq v2.5 flow cell?

The flow cell packaging is updated to reflect the NextSeq v2.5 naming but the flow cell itself looks the same.

Are there any changes to the protocol for preparing NextSeq v2.5 flow cells?

No, the NextSeq v2.5 flow cells follow the same protocol as the NextSeq v2 kits and have been optimized to use the same loading concentration.

Reagents and Flow Cells

  • Are NextSeq flow cells available in single-read and paired-end versions?

    All NextSeq kits include reagents for paired-end sequencing and a paired-end flow cell. However, single-read runs are possible on a paired-end flow cell when the run is set up as a single-read run.

  • How many lanes are on a NextSeq flow cell?

    The NextSeq flow cell contains 4 physical lanes. However, libraries are loaded onto the flow cell from a single reservoir. You can sequence a single library or multiple pooled libraries on the flow cell.

  • How many tiles are imaged on a NextSeq flow cell?

    There are 2 types of flow cells available for the NextSeq system, the high-output flow cell and the mid-output flow cell. Both flow cells contain 4 lanes, but the lanes differ in width resulting in a different number of tiles.

    • Lanes on the high-output flow cell are imaged top and bottom in 3 swaths, or columns, for a total of 864 tiles.
    • Lanes on the mid-output flow cell are imaged top and bottom in 1 swath for a total of 288 tiles.

    For more information, see the NextSeq 500 System Guide or NextSeq 550 System Guide.

  • What kit sizes are available for the NextSeq system?

    NextSeq kits are available in three sizes:

    • The high-output kit is available in300 cycles (2 x 150), 150 cycles (2 x 75), and 75 cycles (1 x 75).
    • The mid-output kit is available in 300 cycles (2 x 150), 150 cycles (2 x 75).

    For output specifications, see NextSeq 500 specifications page

  • Is a wash required before or after scanning a BeadChip?

    No, a wash is not required after a scan. However, if the system has been dry for 7 days, a wash is required before proceeding to another scan.

  • What BeadChips are supported on the NextSeq 550 system?

    Supported BeadChips include the CytoSNP-850K, HumanCytoSNP-12, and HumanKaryomap-12.

Workflow

  • How many BeadChips can I scan at one time on the NextSeq 550 system?

    One BeadChip can be scanned at a time.

  • How often do I need to perform an instrument wash on the NextSeq 550 system?

    The NextSeq software performs an automatic post-run wash on the NextSeq 550 system after a successful sequencing run.

    • If the reagent cartridge and buffer cartridge are in place and the instrument is idle, a quick wash is required every 14 days.
    • If the instrument is in a dry state, for example only performing array scans using the BeadChips, and the reagent cartridge and buffer cartridge are removed, a quick wash is required every 7 days. 

  • How is scanning enabled on the NextSeq 550 system?

    Array scanning is enabled out of the box on the NextSeq 550 system.

  • How long does it take to complete an array scan on the NextSeq 550 system?

    Scanning takes approximately 40 minutes per BeadChip. The CytoSNP-850K scans at a rate of approximately 5 minutes per sample. The HumanCyto-SNP12 and HumanKaryomap-12 scan at a rate of 3.5 minutes per sample.

  • Can I store the BeadChip adapter on the imaging compartment stage when the NextSeq 550 system is not in use?

    Yes; however, do not store the BeadChip adapter without a BeadChip on the imaging compartment stage on the NextSeq 550 system. 

  • How does the array data quality on the NextSeq 550 system compare to an iScan system?

    NextSeq 550 array data quality is comparable to the iScan system. For instance, both the iScan system and NextSeq 550 system can achieve > 99% SNP calls with a > 99% concordance (R^2) for the Infinium CytoSNP-850k BeadChip.

  • Is NaOCl required for the NextSeq instrument wash?

    A dilute solution of NaOCl is required for the automatic post-run wash.The required NaOCl is included in the reagent cartridge provided in the NextSeq 500/550 Kit v2 and the TG NextSeq 500/550 Kit. However, if you are using the original NextSeq 500 Kit (v1), load 3 ml NaOCl in reservoir #28 before loading the reagent cartridge.

    Requirements for NaOCl differ for manual instrument washes depending on the version of control software you are using:

    • The Manual Post-Run Wash option introduced with NCS v1.4 requires diluted NaOCl.
    • The manual wash option provided with NCS v1.3 or earlier versions does not require NaOCl.

  • How long does the cluster generation step take on the NextSeq system?

    Cluster generation is the first step in the sequencing run in which single DNA molecules are bound to the surface of the flow cell, and then amplified to form clusters. Cluster generation on the NextSeq system takes about 2 hours and 20 minutes.

  • How do I start an array scan following a completed sequencing run on the NextSeq 550 system?

    Following the sequencing run, the NCS v2.0 software initiates an automatic post-run wash on the NextSeq 550 system, which takes about 90 minutes. When the wash is complete, the Home button becomes active and you can begin an array scan.

  • How long after completing a sequencing run can I start an array scan on the NextSeq 550 system?

    You can start an array scan when the NextSeq 550 system completes the automatic post-run wash following a sequencing run.

  • What is causing my scan on the NextSeq 550 system to fail registration?

    Make sure that the BeadChip is seated flat on the adapter and that the BeadChip is free of dust. Use canned air or other compressed dusting method to clear the debris.

  • Can I sequence and scan arrays simultaneously on the NextSeq 550 system?

    No, the BeadChip and flow cell fit in the same imaging compartment stage and use the same cameras.

  • How long does it take to perform an instrument wash on the NextSeq system?

    The NextSeq software performs an automatic post-run wash after each successful sequencing run. The automatic post-run wash takes about 90 minutes.

    NCS v1.4 introduces 2 manual washes: the Quick Wash and the Manual Post-Run Wash.

    • Quick Wash—Uses a Tween 20 solution and takes about 20 minutes.
    • Manual Post-Run Wash—Uses a Tween 20 solution with diluted NaOCl and takes about 90 minutes.

    When you initiate a manual wash with NCS v1.3, the software checks that an automatic post-run wash was performed after the last run.

    • If a wash was not performed, the manual wash takes about 90 minutes.
    • If an automatic post-run wash was performed, the manual wash takes 20 minutes.

  • How long after completing an array scan can I start a sequencing run on the NextSeq 550 system?

    You can start a sequencing run when the NextSeq 550 system completes the array scan.

  • How long does it take to complete a sequencing run on the NextSeq system?

    Run duration depends on the number of cycles performed and the type of flow cell used. Run duration is the same for the NextSeq 500 system and the NextSeq 550 system. For more information, see the NextSeq System specifications page.

  • Do NextSeq reagents use the same SBS chemistry as the HiSeq and MiSeq reagents?

    All Illumina sequencers use the Illumina patented reversible-terminator SBS chemistry. The NextSeq system employs the latest evolution in SBS technology: a novel 2-channel SBS method that supports reduced cycle time and data processing time.

  • How does the data quality compare between the NextSeq 550 system and NextSeq 500 system?

    The NextSeq 500 system and NextSeq 550 system offer the same high sequencing data quality. There is no change to output or quality specifications between runs on the different systems.

Software

  • Can I use Sequencing Analysis Viewer (SAV) to view sequencing data from a NextSeq system?

    Yes. Use SAV 1.8.36, or later to view sequencing data from a NextSeq system. You can download SAV for use on a networked computer or use SAV in BaseSpace. SAV 1.8.36 is also compatible with HiSeq and MiSeq run data.

  • How many cycles are required for template generation on the NextSeq system?

    Template generation, the process of identifying defining cluster positions over the entire flow cell surface, is performed during the first 5 cycles of the sequencing run. To detect a cluster during template generation, there must be at least 1 base other than G in the first 5 cycles.

  • Which version of Illumina Experiment Manager (IEM) is compatible with the NextSeq system?

    A sample sheet is not required for a sequencing run unless you are using the system in standalone mode. Use IEM version 1.8.2, or later, to create a sample sheet.

    If you are connected to BaseSpace, use the BaseSpace Prep tab to record library and indexing information, and to specify other run parameters. When runs are set up on the Prep tab, a sample sheet is not required.

  • How many cycles of sequencing are supported on the NextSeq system?

    The NextSeq system can perform up to a 150-cycle paired-end run (2 x 150) using available NextSeq kits. Kits are available in sizes of 300 cycles, 150 cycles, and 75 cycles. Each kit includes additional cycles for index reads.

  • Is a sample sheet required for a sequencing run on the NextSeq system?

    When connected to BaseSpace, a sample sheet is not required. Library and indexing information is entered on the BaseSpace Prep tab before the run, and the information is passed to the NextSeq system. Available run names appear on the instrument screen during the run setup steps.

    However, if the instrument is configured to run in standalone mode (not connected to BaseSpace), use Illumina Experiment Manager (IEM) v1.8.2, or later, to create a sample sheet.

  • What operating system runs on the NextSeq computer?

    The operating system on the NextSeq computer is Windows 7.

  • What is the output file format from a sequencing run on the NextSeq system?

    The NextSeq system generates base call (BCL) files aggregated by lane with a BCL file for each lane, for each cycle. The aggregate file contains the base call and associated quality score for every cluster.

    When using BaseSpace, BCL files are automatically converted to FASTQ files when data transfer is complete. If BCL files are saved to a local server, use bcl2fastq 2.0 to convert base calls from a NextSeq run. The FASTQ converter must be run on a Linux server.

  • What’s new in NextSeq Control Software (NCS) v2.0?

    NCS v2.0 adds support for scanning BeadChip workflows. Scanning BeadChips on a NextSeq 500 system requires a hardware change. 

  • Is it possible to scan a BeadChip on the NextSeq 550 system when a wash is due?

    No. Always perform a wash when a wash is due. It is not possible to proceed to scanning or sequencing until a wash is performed. Instrument washes are required every 7 days, even when the instrument is used for array scans, since the instrument is in a dry state. Regular washes help maintain the instrument fluidics system.

  • What files do I need to scan a BeadChip on the NextSeq 550?

    To perform a scan on the NextSeq 550, you need Decode (*.dmap) files, a manifest  (*.bpm) file, and a cluster (*.egt) file for the BeadChip you are using. The Decode files are unique for each BeadChip barcode. The manifest and cluster files are unique to the BeadChip product type.

  • How much free space on the NextSeq 550 system hard drive is required for array scanning?

    Array scanning requires about 400 MB of free space on the NextSeq 550 hard drive. NCS v2.0 automatically manages disk space.

  • How do I obtain the Decode files required to scan a BeadChip on the NextSeq 550 system?

    Use the Decode File Client Utility to download Decode files directly from Illumina servers using standard HTTP protocol. You can download the Decode File Client and install it on a computer with access to the network location for the Decode file (DMAP) folder.

  • How do I obtain the cluster and manifest files required to scan a BeadChip on the NextSeq 550 system?

    These files are preloaded in the software, and the location of the files is specified on the BeadChip Scan Configuration screen. From the NCS Home screen, select Manage Instrument, System Configuration, and then BeadChip Scan Configuration. Manifest files use the *.bpm file format, and cluster files use the *.egt file format.

    For more information, see the NextSeq 550 System Guide.

  • What is bcl2fastq2?

    The bcl2fastq2 Conversion Software converts base call (BCL) files generated on the NextSeq system. If you plan to use a third-party data analysis solution outside of BaseSpace, configure the system to operate in standalone mode and use the bcl2fastq2 Conversion Software. For more information, see bcl2fastq Conversion Software.

    Only version 2.0, or later, is compatible with NextSeq data.

Analysis

  • How are base calling and quality scoring performed on the NextSeq system?

    Base calling and quality scoring are performed by an updated implementation of Real-Time Analysis (RTA), called RTA v2, which includes important differences from RTA on other Illumina sequencing systems. For example, all processes are performed in memory to maximize processing speed, and configuration files and output file formats are different. For more information, see the NextSeq 500 System Guide or NextSeq 550 System Guide.

    The workflow includes the following steps:

    • Template generation—Defines the position of each cluster on the flow cell.
    • Imaging—In real time, images are passed in memory to RTA v2.
    • Base calling—Determines a base for every cluster at a specific cycle. After a base is called, RTA v2 assigns a quality score. The result is written to base call (BCL) files, which are aggregated in one file per lane per cycle.

  • Does the NextSeq system perform data analysis on the instrument?

    The data set generated by the NextSeq system is too large for on-instrument analysis. Data must be transferred to BaseSpace or a local server for secondary analysis.

  • Do I need a dedicated server to analyze run data from the NextSeq system?

    You can transfer data from a sequencing run to BaseSpace for analysis and storage. Additionally, you can configure the NextSeq system to transfer data to a local server and perform analysis using third-party software.

  • What happens if my data connection is interrupted during a run on the NextSeq system?

    In the event that data transfer is interrupted during a run, data are stored temporarily on the instrument computer until the connection is restored. When the connection is restored, transferring of data resumes automatically.

    If the connection is not restored before the end of the run, data must be removed from the instrument computer manually before a subsequent run can begin.

  • How do I analyze the array data from the NextSeq 550 system?

    The scan output default file format is GTC, which allows for direct analysis using the BlueFuse Multi software. If you prefer to use GenomeStudio for analysis, you can configure the instrument to generate IDAT files for export.