Site Prep/Lab Environment

Site Prep/Lab Environment

  • Instrument Dimensions
  • Power Specifications
  • Lab Bench Guidelines
  • User-Supplied Consumables and Equipment
  • Instrument Dimensions

    The HiSeq X and dedicated workstation have the following dimensions after installation:

    Measurement

    Instrument

    Control Computer

    Width

    118.6 cm (46.7 in.)

    The waste bottle extends an additional
    10 cm (4 in.) on the right side of the instrument.

    21.6 cm (8.5 in.)

    Height

    94 cm (37 in.) Minimum height

    The instrument height is adjustable to an additional 1.27 cm (0.5 in.).

    57 cm (22.3 in.)

    Depth

    76 cm (30 in.)

    The keyboard tray extends an additional
    19 cm (7.5 in.) from the front panel.

    55.4 cm (21.8 in.)

    Weight

    221.4 kg (488 lbs.)

    34 kg (75 lbs.)

    Power Specifications

    Type

    Specification

    Line Voltage

    100–240 Volts AC @ 50/60 Hz

    Power Consumption

    Maximum 1500 Watts

    Lab Bench Guidelines

    Bench-Craft (www.bench-craft.com), part # HS-30-60-30 P2 with casters.

    Width

    Height

    Depth

    Locking Casters

    152.4 cm (60 in.)

    76.2–91.4 cm (30–36 in.)

    76.2 cm (30 in.)

    Yes

    User-Supplied Consumables and Equipment

    For a complete list of user-supplied consumables, user-supplied equipment, and consumable storage requirements, see the HiSeq X Lab Setup and Site Prep Guide.