Which bisulfite conversion kit should I use for the Infinium Methylation arrays?

03/29/2022


This bulletin is intended for customers using Infinium Methylation BeadChips. Before carrying out the Infinium HD Methylation Assay, samples must first be bisulfite converted. The efficiency and accuracy of the bisulfite conversion step, including using the final elution volume recommended by Illumina, is critical to the accuracy and performance of the Infinium Methylation assay.

Illumina has validated and recommends only the following kits from Zymo Research.

EZ DNA / EZ-Methylation Kits

Bisulfite conversion completed in ~ 1.5 days

Product Name

Size

Format

Catalog No.

Elution Volume
500 ng /1000 ng

Preparation

EZ-96 DNA Methylation™ Kit

2 x 96 rxn

96-well- (Deep)

D5004

12 µl/22 µl

Manual

EZ DNA Methylation™ Kit

50 rxn

Spin-Column

D5001

12 µl/ 22 µl

Manual

EZ DNA Methylation™ Kit

200 rxn

Spin-Column

D5002

12 µl/ 22 µl

Manual

 

 

 

 

 

 

 

EZ-96 DNA Methylation Lightning MagPrep Kits (uses magnetic beads for purification)

Bisulfite conversion completed in ~ 3 hours

Product Name

Size

Format

Catalog No.

Elution Volume

Preparation

EZ-96

DNA Methylation-Lightning MagPrep Kit

4 x 96 rxn

96-well (Deep)

D5046

25 µl

Manual

EZ-96

DNA Methylation-Lightning MagPrep Kit

8 x 96 rxn

96-well (Deep)

D5047

25 µl

Manual

EZ DNA Methylation-Lightning Automation Kit

1 x 96 rxn

96-well (Deep)*

D5049

25 µl

Automated**

 

 

 

 

 

 

 

 

*User provides 96-well deep-well microplates and PCR plates

**The EZ-96 DNA Methylation-Lightning Automation Kit is equivalent to the EZ-96 DNA Methylation-Lightning MagPrep Kit but contains buffer and reagent volumes needed for procedures using liquid-handling automation. Liquid-handling automation scripts and setup support are available through Zymo Research. For questions regarding automation, contact automation@zymoresearch.com

 

Not recommended:

  • EZ-96 DNA Methylation™ Kit (Shallow Well Format), Zymo Research Catalog Number D5003. (The final step of the bisulfite conversion before beginning the Infinium HD assay is the elution of the converted sample from the spin column. The Shallow Well format of the EZ-96 DNA Methylation Kit is not recommended for use with the Infinium Methylation HD Assay because the elution volume of 30 µl is too large for inclusion in the Infinium HD Assay.)
  • The following kits have not been validated and are not recommended for use in preparing samples for the Infinium Methylation BeadChip
    • EZ DNA / EZ-96 DNA Methylation-Gold Kits
    • EZ DNA / EZ-96 DNA Methylation-Direct Kits
    • EZ DNA / EZ-96 DNA Methylation-Lightning Kits
      • Spin Column: D5030
      • 96-well (Shallow): D5032
      • 96-well (Deep): D5033 
  • No other bisulfite conversion kits have been validated.

 